The Staglin EEG Suite is part of the [http://www.semel.ucla.edu/staglin Staglin IMHRO Center for Cognitive Neuroscience] and contains resources including:
This page will give you general information about the EEG Suite including how to use the testing rooms, EGI net information, current EEG projects, and where to locate and order supplies.
*[http://www.egi.com/research-division-research-products/eeg-systems/251-conventional-eeg EGI Net Amps 300]
*[http://www.egi.com/research-division-research-products/eeg-systems/191-ges300mr MR-safe EGI Net Amps 300]
*[http://www.egi.com/research-division-research-products/sensor-registration EGI Photogrammetry System]
*[http://www.egi.com/research-division-research-products/sensor-nets EGI HydroCel EEG Nets] - S/M/L of MR and non-MR compatible versions


Follow the steps below when collecting EEG data at the Staglin EEG Suite.


'''We recommend giving these directions a complete once-over before you ever take data here.'''
==EEG Facilities & Equipment==
The Staglin EEG Suite is part of the [http://www.semel.ucla.edu/staglin Staglin IMHRO Center for Cognitive Neuroscience] and contains resources including two fully shielded EEG suites and a prep room.


The Left EEG Suite:
*[https://www.egi.com/images/stories/manuals/GES_300/GES_300_tman_8103003-51_20150304_hires.pdf EGI Net Amps 300]
*[https://www.egi.com/images/stories/placards/legacy/PIB_instructions_plac_8404162-51_20100427.pdf Polygraph Input Box (PIB) for Net Amps 300]
*[https://www.egi.com/images/stories/company/documents/net_amps_400_flyer_oct_2015_MM_421.pdf MR-safe EGI Net Amps 400]
*[https://www.egi.com/research-division/geodesic-sensor-net EGI HydroCel EEG Nets] - S/M/L of MR and non-MR compatible versions


The Right EEG Suite:
*Bio-Semi 64 channel EEG system using active electrodes


==Step 1: Make a Reservation==
==Using the EEG Suite==
There is a [https://www.google.com/calendar/embed?src=28vln94hoc2da1k40rn6lcp1bs%40group.calendar.google.com&ctz=America/Los_Angeles Google Calendar] for reserving the Staglin EEG Suite.
This page describes step-by-step how to schedule, set-up, and test in the Staglin EEG Suite:   
 
*[[StaglinCenter:EEG:Testing]]
Request access to this calendar by emailing Agatha (alenarto @ ucla.edu).  When you are ready to take EEG data, make a reservation on this calendar ahead of time.
 
==Step 2: Setup for Data Collection==
 
===Before the Subject Gets There===
====Booting the Computer Properly====
We maintain two different versions of the NetStation program for data collection.  Follow these steps to boot the computer properly to use the version you want.  '''Remember, consistency in your data collection is important, so the version of NetStation matters.'''
*[[StaglinCenter:EEG:BootVolume]]
 
====A/V Setup====
The data collection room is setup so you can pipe in audio and visual stimuli to your subject.  Follow these steps to set this up properly.
*[[StaglinCenter:EEG:A/V]]
 
====Capturing Subject Responses====
If your experimental paradigm includes taking subject responses, read this to know how to record them.
*[[StaglinCenter:EEG:Responses]]
 
====Event Timing====
If your experimental paradigm records when responses and stimuli occur, we has a system for accurately timing these events and recording them with the EEG signal.
*[[StaglinCenter:EEG:Timing]]
 
===Once the Subject Arrives===
====EEG Net Prep and Application====
EGI has some how-to's on this, and we recommend you read the manuals (located in the cabinet above the table in the EEG collection room).  But we've got a step-by-step here for you.
*[[StaglinCenter:EEG:Net]]
 
====Photogrammetry====
If you care to use the [http://www.egi.com/research-division-research-products/sensor-registration EGI Photogrammetry System] to record the electrode placement on your subject. This is useful for [[EEG:Processing:Source|source localization]].
*[[StaglinCenter:EEG:Photogrammetry]]


'''We recommend giving these directions a complete once-over before you ever take data here.'''


==Step 3: Collect Data==
===Starting Net Station===
Once the net has been placed on your subject, follow these steps to properly start up Net Station.
*[[StaglinCenter:EEG:StartingNetStation]]
===Impedance Check and Fix===
You need to check the impedance of the electrodes and fix any that are too high or else your data collection will not be ideal.
*[[StaglinCenter:EEG:ImpedanceCheck]]
===Collection===
Where to save your data so you are a responsible user and can find it in the future.
*[[StaglinCenter:EEG:Collect]]
==Step 4: Feedback Form==
For quality control, we devised a feedback form that we ask researchers to fill out each time they collect data in the Staglin EEG Suite.  Follow the instructions written on the Desktop Background to access and fill out this form. 
The identification names of each net can be found on the plug.
We actively monitor the feedback in this form.  If you'd like to know see this feedback, email Edward at '''eplau @ ucla.edu''' .
==Step 5: Cleaning Up==
===Net Disinfection===
After each use, rinse out and disinfect the net by following this step-by-step process.
*[[StaglinCenter:EEG:Disinfect]]


==Troubleshooting and Denoising==
==Troubleshooting and Denoising==
*[[Checking Ground]]
*[[Checking Ground]]
*[[Checking for wire loops]]
*[[Checking for Wire Loops]]
*[[Checking Scanner Settings]]
*[[Checking Scanner Settings]]
*[[Checking Cap Electrodes for Noise]]
*[[Checking Cap Electrodes for Noise]]


==EEG Suite supplies==
===Where to find EEG supplies in the lab===
Most of the supplies needed for EEG testing is located in the EEG net room in the wooden cabinet by the sink.  If you open the left cabinet, on the main top shelf, most of the supplies you will need for testing will be located in the plastic bin labeled with EEG supplies.  You will find the following in that bin:
*'''scissors'''
*'''red wax pencil'''
*'''tape measure'''
*'''spandages'''
*'''disposable EEG/MR electrodes'''


'''Johnson's baby shampoo''', '''EEG solution salt''', '''extra electrodes''', '''the EGI net manual''', and '''EGI net boxes'''  along with their respective measuring spoons are usually located on top of the cabinet by the sink. The electrodes and the net boxes are in the cardboard boxes on top of the cabinet.
==EGI Net Maintenance==
*[https://www.egi.com/images/stories/manuals/Printed%20IFUs%20with%20New%20Notified%20Body/HC_GSN_RM_sponged_appl_uins_8403481-53_20181210.pdf Procedure for Routine Net Tests]
*[https://docs.google.com/forms/d/1qYeNEjPSWKaGmElK9eYHlLOpIOF6R5L1tYh1l861R4E/viewform EGI Net Check-Up Form]
*[https://docs.google.com/spreadsheets/d/1Gx4vqWxGfCX9JPs5AoBQcIU4OveudAx9vv4XIcNJ8YQ/edit#gid=332429759 EGI Net Check-Up Log]


In the EEG Suite's left testing room, you will find remaining supplies in the cabinets to the left of the table with the computer monitor and the EGI amp.  The cabinet shelves are labeled with these items:
*'''latex gloves'''
*'''disinfectant'''
*'''pipettes'''
*'''medical tape'''
*'''EEG salt'''


If you see that something has run out, check the following places to see if extras are stored:
==EEG Suite supplies==
*the wooden cabinet by the sink the EEG net room (Johnson's baby shampoo and spandages may be hiding in the shelf below the plastic bins, in the bottom drawers, or on top of the cabinet)
*[[Where to find EEG supplies]]
*the cabinets in the EEG suite (currently, there are MANY boxes in the right-most cabinet in the testing room and some on the top shelf of the cabinet where the Stimulus computer lives).
*[[Where to order EEG supplies]]
<!---*[[Current supply stock]]--->


'''NOTE:'''  If something has completely run out, there is an emergency supply located in the shop, but please ask Agatha, Sam, or Cameron for these supplies.  BUT ONLY FOR EMERGENCIES.


===Where to order EEG supplies===
==Current Projects==
We have had success ordering supplies from the following places.  The ones ordered from online sites accept purchase orders:
===Concurrent EEG/fMRI===
*'''pipettes''' (5.8mL, non-sterile):  [http://www.fishersci.com/ecomm/servlet/fsproductdetail?storeId=10652&productId=634785&catalogId=29104&matchedCatNo=1371120||137119AM||137119CM||137119BM&fromSearch=1&searchKey=pipette||pipettes||pipeters||pippette||pipettor||pipetter||pipeter||pipettors||pipetters&highlightProductsItemsFlag=Y&endecaSearchQuery=%23store%3DScientific%23nav%3D0%23rpp%3D25%23offSet%3D0%23keyWord%3Dpipettes%23searchType%3DPROD%23SWKeyList%3D Fisher Scientific]
*[[SWM Concurrent]]
*'''MT Spandages''' (MT07, 7-inch spandage, 25 yards):  [http://www.vitalitymedical.com/mt-spandage-tubular-elastic-dressing.html Vitality Medical MT Spandage] 
*'''Johnson's baby shampoo''':  This is very hard to find online, so purchase from a store (CVS, Target, Walgreens, etc.) and give the bank statement and receipt to Jun or Malina for a reimbursement.
*'''EGI KCL salt''':  order from EGI by e-mail supportteam@egi.com
*'''Control3 disinfectant''':  e-mail control3@earthlink.net - order the 16 oz. bottle.
*'''GSR electrodes''' (EL509, 100 electrodes/box):  [http://www.biopac.com/disposable-radiotranslucent-dry-electrode Biopac Systems, Inc.]
*'''Self-Adhesive Bandage''' (Item 2082, 2 inch x 5 yards, latex-free):  [http://www.vitalitymedical.com/coban-self-adherent-wrap.html Vitality Medical Self-Adhesive Wrap]
 
===Current supply stock===
*'''pipettes''':  ordered 2014.01.10 - 10 boxes of 500 pipettes
*'''Johnson's baby shampoo''':  ordered 2014.01.20 - 8 bottles
*'''MT Spandages''':  ordered 2014.08.06 - 10 boxes
*'''GSR electrodes''':  ordered 2014.2.12 - 2 boxes of 100 electrodes
*'''Control III Disinfectant''':  ordered 2014.4.4 - 3 16 oz. bottles
*'''EGI KCL Salt''':  ordered 2014.3.27 - 5 tubs
*'''Self-Adhesive Bandage''':  ordered 2014.3.15 - 36 rolls
 
==Current EEG Projects==
*[[StaglinCenter:EEG:SSVEP]]




Line 129: Line 49:
If you want to take EKG data concurrently...
If you want to take EKG data concurrently...
*[[StaglinCenter:EEG:EKG]]
*[[StaglinCenter:EEG:EKG]]




==External Links==
==External Links==
*[http://www.egi.com EGI]
*[http://www.egi.com EGI]
*[http://www.egi.com/research-division-research-products/eeg-systems/251-conventional-eeg EGI Amp]
*[https://www.egi.com/research-division/net-amps-eeg-amplifier EGI Amp]
*[http://www.egi.com/research-division-research-products/eeg-systems/191-ges300mr EGI MR Safe Amp]
*[https://www.egi.com/images/stories/manuals/GES_400/core/GES_400MR_uman_8100401-55_20150722_hires.pdf EGI MR Safe Amp]
*[http://www.egi.com/research-division-research-products/sensor-registration EGI Photogrammetry System]
*[https://www.egi.com/knowledge-center EGI Knowledge Center]

Latest revision as of 23:18, 26 September 2021

This page will give you general information about the EEG Suite including how to use the testing rooms, EGI net information, current EEG projects, and where to locate and order supplies.


EEG Facilities & Equipment

The Staglin EEG Suite is part of the Staglin IMHRO Center for Cognitive Neuroscience and contains resources including two fully shielded EEG suites and a prep room.

The Left EEG Suite:

The Right EEG Suite:

  • Bio-Semi 64 channel EEG system using active electrodes

Using the EEG Suite

This page describes step-by-step how to schedule, set-up, and test in the Staglin EEG Suite:

We recommend giving these directions a complete once-over before you ever take data here.


Troubleshooting and Denoising


EGI Net Maintenance


EEG Suite supplies


Current Projects

Concurrent EEG/fMRI


Related Info

EKG

If you want to take EKG data concurrently...


External Links