Eye Tracking: Difference between revisions

From Center for Cognitive Neuroscience
Jump to navigation Jump to search
Line 14: Line 14:
**One skinny, orange cable connected to a blue cable: Fiber optic cable. Plug it into its port on the back of the eyetracker.
**One skinny, orange cable connected to a blue cable: Fiber optic cable. Plug it into its port on the back of the eyetracker.
*Remove the lens cap. Place it somewhere nearby so you can find and replace it again easily when your scan is complete.
*Remove the lens cap. Place it somewhere nearby so you can find and replace it again easily when your scan is complete.
*At this point, the Eyelink desktop computer in the control room can be turned on. The camera now needs to be aligned so that it points at the participant's eye (usually right) reflected in the mirror.
*At this point, the Eyelink desktop computer in the control room can be turned on. The camera now needs to be aligned so that it points at the participant's (usually right) eye, reflected in the mirror.
*Adjust the alignment as necessary. You may need to play around with both wheels until a stable image is achieved:
**Remember to use the mirror dedicated to eyetracking, not the usual coil mirrors. The eyetracking mirror is stored in the control room cabinet and labelled with EYETRACKER.
*Once the participant is inside the bore, adjust the alignment as necessary. You may need to play around with both wheels until a stable image is achieved:
**Wheel close to lens: Focus wheel.
**Wheel close to lens: Focus wheel.
**Wheel close to base: Aperture wheel.
**Wheel close to base: Aperture wheel.
**Large knob near the base: Adjust eyetracker positioning in all directions.
**Large knob near the base: Adjust eyetracker positioning in all directions.
==In Control Room==
==In Control Room==
*At the Eyelink station on the right side of the room, there is a cable with a yellow tag on it and connected to a battery box labelled EYELINK (pictured below). This is the transducer cable--plug it into the rightmost port (labelled with "12V") on the box on top of the tower unit.
*At the Eyelink station on the right side of the room, there is a cable with a yellow tag on it and attached to a battery box labelled EYELINK (pictured below). This is the transducer cable--plug it into the rightmost port (labelled with "12V") on the box on top of the tower unit.
[[File:EYELINK_transducer.jpg]]
[[File:EYELINK_transducer.jpg|250px|thumb|right|Transducer cable.]]
*Once the cables are connected to the eyetracker in the bore, turn on the tower unit. You will likely get errors if you try to power on before those cables are connected.
*Once the cables are connected to the eyetracker in the bore, turn on the tower unit. You will likely get errors if you try to power on before those cables are connected.
*The main Eyelink menu screen should appear, now that all the cables are connected and the computer is powered on.
==Calibration==
*Click on the center of the eye to put the red circle around the pupil.
*Set threshold:
**Start by clicking "Auto Threshold". Eyelink will give you its best estimate of what threshold values give you the most accurate capture of pupil and corneal reflection.

Revision as of 17:49, 29 July 2022

EyeLink

MR Compatible EyeLink 1000

Specifications

Image-size-projector-screen.png

Eyelink-in-scanner-eye-to-screen-distance.png

Setup

In Scanner

  • The eyetracker is installed in the back of the bore. It should stay behind the blue tape marking the best position on the rail.
  • There are three cables that need to be attached when you begin an eyetracking session:
    • Two thick, black cables: Power cables. Plug them into the two round ports on the back of the eyetracker (doesn't matter which one goes on which side).
    • One skinny, orange cable connected to a blue cable: Fiber optic cable. Plug it into its port on the back of the eyetracker.
  • Remove the lens cap. Place it somewhere nearby so you can find and replace it again easily when your scan is complete.
  • At this point, the Eyelink desktop computer in the control room can be turned on. The camera now needs to be aligned so that it points at the participant's (usually right) eye, reflected in the mirror.
    • Remember to use the mirror dedicated to eyetracking, not the usual coil mirrors. The eyetracking mirror is stored in the control room cabinet and labelled with EYETRACKER.
  • Once the participant is inside the bore, adjust the alignment as necessary. You may need to play around with both wheels until a stable image is achieved:
    • Wheel close to lens: Focus wheel.
    • Wheel close to base: Aperture wheel.
    • Large knob near the base: Adjust eyetracker positioning in all directions.

In Control Room

  • At the Eyelink station on the right side of the room, there is a cable with a yellow tag on it and attached to a battery box labelled EYELINK (pictured below). This is the transducer cable--plug it into the rightmost port (labelled with "12V") on the box on top of the tower unit.
Transducer cable.
  • Once the cables are connected to the eyetracker in the bore, turn on the tower unit. You will likely get errors if you try to power on before those cables are connected.
  • The main Eyelink menu screen should appear, now that all the cables are connected and the computer is powered on.

Calibration

  • Click on the center of the eye to put the red circle around the pupil.
  • Set threshold:
    • Start by clicking "Auto Threshold". Eyelink will give you its best estimate of what threshold values give you the most accurate capture of pupil and corneal reflection.